Dialyzer Reprocessing Machine KDR-A200

Dialyzer Reprocessing Machine KDR-A200

Dialyzer Reprocessing Machine KDR-200 is an advanced dialyzer reprocessing unit equipped with touch screen interface to navigate, monitor and select various functional parameters and settings. Specialized inflow pressure sensor and blood volume sensor enables synchronous workflow. Comprehensive storage of time-stamped dialyzer data improves standardization and stores essential determinants for further use of dialyzer.

  • Cycle time : 8 minutes per cycle
  • Volume test range : 0 mL to 150 mL ( ±3 % )
  • Operational Workstation :

Specification

Cycle time 8 minutes per cycle
Volume test range 0 mL to 150 mL ( ±3 % )
Maximum water inlet flow > 1600 mL/min
Water consumption 8 L
Water input temperature 10 °C to 30 °C
Water input pressure 0.15 MPa to 0.35 Mpa
Pressure lost test Low-flux dialyzer <0.83 mmHg/s
High-flux dialyzer <1.25 mmHg/s
Standard rinse 11 min for one or 22 min for two
Intensive rinse 14 min for one or 27 min for two
Power supply 220 V / 50 Hz
Dimensions 480 × 430 × 560 mm
Weight 25 Kgs

Features

  • Automated reprocessing cycle: 8 minutes per cycle

  • Automatic self-test

  • Touch screen LCD

  • Rinse : Intensive and Standard

  • Data record and back up

  • Alarms for tube clogging

  • Auto-sensing pressure and blood volume

  • Reduced water loss

Applications

Used in nephrology (AKD) departments in hospitals or Kidney dialysis centers to effectively automate reprocessing activities detect blood volume as well as inspect for dialyzer compliance with hospital infection-control measures for reuse of the dialyzer unit and provide recorded patient data for record keeping and quality control.